ELECTROCHEMICAL CORROSION ANALYZER

 

Name of the equipment:

Electrochemical Corrosion Analyzer

 

Make & Model:

CH Instruments Inc., USA

CHI608E Potentiostat/Galvanostat

 

I-Stem Registration ID-

3224881                            

 

Category of Instrument

Electrochemical Instrument

 

Types of Analysis / Testing

Corrosion Testing

 

Application:

Studying the corrosion resistance of various metals and alloys.

Description of Instrument

Booking DetailsA Corrosion testing equipment.

 

 

 

 

Book through I-STEM:

https://www.istem.gov.in/

 

Slot Booking Link

I-STEM Slot Booking link for External User

Booking available for

Internal and External Both

 

Requisition form for

Internal

External

 

Contact Details

 

 

Faculty In-charge:

Dr. Murali KP

 

Email ID:

kpmurali@nitc.ac.in

 

 

Phone number:

 0495-2286414

Technical Staff:

Sanal PR (TA SG I)  

sanal@nitc.ac.in

 

 

Jithin Krishna M (TA)

jithinkrishnam@nitc.ac.in

 

Jijeesh k (ST)

jijeeshk@nitc.ac.in

 

Features, Working Principle and Specifications

Department

MED

 

Office Email ID

medoffice@nitc.ac.in

 

Location

Material Science Lab

Production Block

 

Lab Phone No

0495-2286450

 

 

 

 

 

Potentiostat

 

·         Zero resistance ammeter

·         2- or 3- or 4-electrode configuration

·         Floating (isolated from earth) or earth ground

·         Maximum potential: ±10 V

·         Maximum current: ±250 mA continuous, ±350 mA peak

·         Compliance Voltage: ±13 V

·         Potentiostat rise time: < 1 μs, 0.8 μs typical

·         Applied potential ranges (volts): ±0.01, ±0.05, ±0.1, ±0.65, ±3.276, ±6.553, ±10

·         Applied potential resolution: 0.0015% of potential range

·         Applied potential accuracy: ±1 mV, ±0.01% of scale

·         Applied potential noise: < 10 μV rms

·         Measured current range: ±10 pA to ±0.25 A in 12 ranges

·         Measured current resolution: 0.0015% of current range, minimum 0.3 fA

·         Current measurement accuracy: 0.2% if current range >=1e-6 A/V, 1% otherwise

·         Input bias current: < 20 pA

 

Galvanostat

 

·         Galvanostat applied current range: 3 nA - 250 mA

·         Applied current accuracy: 20 pA ±0.2% if > 3e-7A, ±1% otherwise

·         Applied current resolution: 0.03% of applied current range

·         Measured potential range (volts): ±0.025, ±0.1, ±0.25, ±1, ±2.5, ±10

·         Measured potential resolution: 0.0015% of measured range

Type of Sample Required for Analysis / Testing (Quantity, Pre-Preparation, State etc.) Guidelines for Sample Submission – User Instructions

Galvanostat

Electrometer

 

·         Reference electrode input impedance: 1x1012 ohm

·         Reference electrode input bandwidth: 10 MHz

·         Reference electrode input bias current: <= 10 pA @25°C

 

 

Waveform Generation and Data Acquisition

·         Fast waveform update: 10 MHz @ 16-bit

·         Fast data acquisition: dual channel 16-bit ADC, 1,000,000 samples/sec simultaneously

·         External signal recording channel at maximum 1 MHz sampling rate

 

 

  Experimental Parameters

 

·         CV and LSV scan rate: 0.000001 to 10,000 V/s

·         Potential increment during scan: 0.1 mV @ 1,000 V/s

·         CA and CC pulse width: 0.0001 to 1000 sec

·         CA and CC minimum sample interval: 1 μs

·         True integrator for CC

·         DPV and NPV pulse width: 0.001 to 10 sec

·         SWV frequency: 1 Hz to 100 kHz

·         i-t sample interval: minimum 1 μs

·         ACV frequency: 0.1 Hz to 10 kHz

·         SHACV frequency: 0.1 Hz to 5 kHz

·         FTACV frequency: 0.1 Hz to 50 Hz, simultaneously acquire 1st, 2nd, 3rd, 4th, 5th, and 6th harmonics ACV data

·         IMP frequency: 0.00001 Hz to 1 MHz

·         IMP amplitude: 0.00001 V to 0.7 V rms

 

 

 

 

 

 

 

 

 

·         Sample Type: Metals, Alloys and Conductive materials

·         Pre-preparation: The sample should be clean & polished.

·         Dimensions for the sample: Should be a flat surface that must cover a circular hole with diameter 10 mm and thickness must not exceed 10 mm and should have a minimum thickness of 1 mm.

·         Type of Samples to be Analysed - Metallic

·         Maximum No. of Samples Accepted at a Time- 1

·         Minimum No of Days Required for Analysis – 2 Days

 

User Charges Rs. (GST Extra)

 

 

Internal

External Academic Institutes

National R&D Lab

Industry

200/- per hour

500/- per hour + GST

500/- per hour + GST

500/- per hour + GST