Potentiostat
·
Zero
resistance ammeter
·
2-
or 3- or 4-electrode configuration
·
Floating
(isolated from earth) or earth ground
·
Maximum
potential: ±10 V
·
Maximum
current: ±250 mA continuous, ±350 mA peak
·
Compliance
Voltage: ±13 V
·
Potentiostat
rise time: < 1 μs, 0.8 μs typical
·
Applied
potential ranges (volts): ±0.01, ±0.05, ±0.1, ±0.65, ±3.276, ±6.553, ±10
·
Applied
potential resolution: 0.0015% of potential range
·
Applied
potential accuracy: ±1 mV, ±0.01% of scale
·
Applied
potential noise: < 10 μV rms
·
Measured
current range: ±10 pA to ±0.25 A in 12 ranges
·
Measured
current resolution: 0.0015% of current range, minimum 0.3 fA
·
Current
measurement accuracy: 0.2% if current range >=1e-6 A/V, 1% otherwise
·
Input
bias current: < 20 pA
Galvanostat
·
Galvanostat
applied current range: 3 nA - 250 mA
·
Applied
current accuracy: 20 pA ±0.2% if > 3e-7A, ±1% otherwise
·
Applied
current resolution: 0.03% of applied current range
·
Measured
potential range (volts): ±0.025, ±0.1, ±0.25, ±1, ±2.5, ±10
·
Measured
potential resolution: 0.0015% of measured range

|
Galvanostat
Electrometer
·
Reference
electrode input impedance: 1x1012 ohm
·
Reference
electrode input bandwidth: 10 MHz
·
Reference
electrode input bias current: <= 10 pA @25°C
Waveform Generation
and Data Acquisition
·
Fast
waveform update: 10 MHz @ 16-bit
·
Fast
data acquisition: dual channel 16-bit ADC, 1,000,000 samples/sec
simultaneously
·
External
signal recording channel at maximum 1 MHz sampling rate
Experimental Parameters
·
CV
and LSV scan rate: 0.000001 to 10,000 V/s
·
Potential
increment during scan: 0.1 mV @ 1,000 V/s
·
CA
and CC pulse width: 0.0001 to 1000 sec
·
CA
and CC minimum sample interval: 1 μs
·
True
integrator for CC
·
DPV
and NPV pulse width: 0.001 to 10 sec
·
SWV
frequency: 1 Hz to 100 kHz
·
i-t
sample interval: minimum 1 μs
·
ACV
frequency: 0.1 Hz to 10 kHz
·
SHACV
frequency: 0.1 Hz to 5 kHz
·
FTACV
frequency: 0.1 Hz to 50 Hz, simultaneously acquire 1st, 2nd, 3rd, 4th, 5th,
and 6th harmonics ACV data
·
IMP
frequency: 0.00001 Hz to 1 MHz
·
IMP
amplitude: 0.00001 V to 0.7 V rms
|