3D bioprinter

 

Name of the equipment:

3D bioprinter

 

Make & Model:

Cellink ; BioX Gen III

 

I-Stem Registration ID-

3246253

 

Category of Instrument

Bioscience and Technology

 

Types of Analysis / Testing

Fabrication

 

Application:

Bioprinting scaffolds for disease models, drug testing, etc.

Description of Instrument

A 3D bio-printer for cell biologists, tissue engineers and biomedical researchers

Booking Details

 



Book through I-STEM:

https://www.istem.gov.in/

 

Slot Booking Link

I-STEM Slot Booking link for External User

Contact Details



Booking available for

Internal and External Both

 

Requisition form for

Internals

Externals

 

 

 

 

Faculty In-charge:

Dr. Baiju G Nair

Email ID:

bgnair@nitc.ac.in

Phone number:

 +91-7034011575

Technical Staff:

Sandhyadevi S M (TA) sandhyadevism@nitc.ac.in

 

Sooraj V S (ST

soorajvs@nitc.ac.in

 

Jishnu Sudhakaran (TE) jishnusudhakaran@nitc.ac.in

 

Sruthi K Balan (LA)

sruthikbalan@nitc.ac.in

 

 

Department

BSED

 

Office Email ID

sbtoffice@nitc.ac.in

 

Location

Room No. 307D, Department of Bioscience and Engineering

 

Lab Phone No

04952285490

 

 

Features, Working Principle and Specifications  

 

 


Features of the equipment

 

ü  Versatile support for 3 print head

ü  Patented clean chamber technology for contamination free 3D bioprinting

ü  Universal bioink capability

 

 

Unique features/Measurement capabilities, if any

Ø  Theoretical 3D printing resolution: 1µm (practical limit 100 µm)

Ø  Integrated DNA studio software for design of simple structures

Ø  Application: Bioprinting scaffolds for disease models, drug testing, etc.

 

Instrument Technical Description and Major Specifications (This Specifications Limited to Major 5)

 

§  Build Volume: 130mm x 90 mm x 70 mm

§  Pressure Range: 0-200 kPa

§  Temperature: (i) Printbed: 4-65oC

§   (ii)Printhead: Amb-250 oC

§  UV-Sterilization: UV-C (275nm) 20mW output

  • Supported File Formats: .gcode, .sti

Measurement/Sample specifications:

§  Viscosity: 1000-10000 mPa.S

§  Only biocompatible polymers allowed | No microbial samples allowed

§  Both biopolymers and synthetic polymers (Melting point: 200°C)



Type of Sample Required for Analysis / Testing (Quantity, Pre-Preparation, State etc.) Guidelines for Sample Submission – User Instructions

 

 





·         Quantity: 5ml (depends on the print model)

·         Sample Type: Viscous fluid (500-8000mPa.S)

·         Pre-preparation: The sample should be sterile

·         User instructions: Should reveal the cell used in the ink

 

User Charges Rs. (GST Extra) 

 


Internal

External Academic Institutes

National R&D Lab

Industry

500/- per hour

1500/- per hour

1500/- per hour

2000/- per hour